Now with an
adjustable alert, the Inspector+ is a small,
handheld, microprocessor-based instrument which offers
excellent sensitivity to low levels of alpha, beta,
gamma, and x-rays. The digital readout is displayed with
a red count light and a beeper sounds with each count
detected. Other features include an adjustable timer,
external calibration controls and selectable alert.
Specifications
Detector Internal
Halogen-quenched, uncompensated GM tube with thin mica
window, 1.4-2.0 mg/cm2 areal density. Effective diameter
of window is 45 mm (1.75 in.). Radiation symbol on front
label marks the center of the detector.
Display 4 digit liquid crystal
display with indicators. Averaging
Periods Display will update every 3 seconds.
At low background levels, the update is the moving
average for the past 30-second time period. The timed
period for the moving average decreases as the radiation
level increases
| |
 |
Gamma
Sensitivity 3500 CPM/mR/hr
referenced to Cs-137. Smallest detectable level
for I-125 is .02 µCi at contact. Timer can set 1
minute sampling periods from 1 to 10
minutes. |
| Operating
Range |
| mR/hr - .001 (1µR) to 100
mR/hr |
CPM - 0 to 350,000 |
| µSv/hr - .01 to 1000 |
CPS - 0 to 5000 |
| Total/ Timer - 1 to 9,999,000
counts |
Accuracy mR/hr
±10% typical (NIST), ±15% max - 0 to
100 µSv/hr ±10% typical (NIST), ±15% max - 01
- 1000 CPM ±10% typical (NIST), ±15% max - 0
to 350,000 (Referenced to
Cs-137) | | |
Alert Set
Range mR/hr .001 - 50 and CPM 1 -
160,000 Count Light Red LED
flashes with each radiation
event. Audio
Indicator Internally mounted beeper
(can be switched off for silent
operation) Outputs Dual
miniature jack drives CMOS or TTL devices: count
to computer or datalogger. Submini jack input
allows for electronic
calibration Anti-Saturation Meter
will hold at full scale in fields as high as 100
times the maximum reading. Power
Requirements One 9-volt alkaline
battery. Battery life is 2160 hours at normal
background. Temperature
Range -10ºC to 50ºC (14ºF to
122ºF) Weight Inspector:
275 grams (9.7 oz.) without
battery. Size 150 x 80 x 30
mm (5.9 x 3.3 x 1.2
in.). Includes Carrying
Case, CE
Certification Warranty 1
year limited, 90 day Warranty on GM Detector |
 |
 |
Utility Menu The
Utility Menu offers several options accessible by
holding down the + button while turning the
instrument on. • 3 second (fast
response) • Switches from mR/hr and CPM to
µSv/hr and CPS • Reset the cal factor to
100. • Probe adjustment settings. • Adjust
the CAL factor. • Reset all setting to default
factory setting. |
Back of
Unit |
| |